Page 19

Teledyne API Model T201 NH3 Analyzer Operator Manual

Calibration Procedure

 

 

Table 2-2. Zero Calibration Procedure - Zero Gas through ZERO Port

 

 

 

 

 

 

 

Step No.

Action

Comment

 

 

 

 

 

 

 

1.

Press CALZ

The T201 enters the calibrate mode from sample

 

 

 

mode. The zero gas is supplied through the ZERO

 

 

 

inlet port on the rear panel of the M501 NH3.

 

2

Table 2.1

Follow Steps 2 to 8 in Table 2-1.

 

 

 

 

 

 

Notes

Since the zero gas concentration is defined as 0 ppb, it is not necessary

 

 

to enter the expected concentration values.

 

 

Both the TNx and NOx channels must be zeroed individually. When you zero the NOx channel both the offsets for NOx and NO are configured. All readings should display zero following this action.

It is recommended that the calibration of both the TNx and NOx channels be done at one time.

Always confirm that excess zero air is flowing past the sample or zero port, if the Zero/Span option is present, on the rear of the M501 NH3 converter. There should always be some excess flow. If insufficient flow is supplied to the analyzer ambient gas will be entrained and the zero offset values will be incorrect. However, it is important not to over pressurize the analyzer.

It is a good idea to allow the analyzer to sample zero gas for extended periods following zero calibration. Large drifts can signify leaks or indicate an issue with the zero air source.

2.2SPANNING THE ANALYZER WITH NITRIC OXIDE GAS

The analyzer can be spanned by either applying nitric oxide gas straight to the Sample port or through the Zero/Span valve option, if installed. It is important to remember that if the analyzer was previously sampling ammonia gas, prior to conducting a nitric oxide span calibration, it will take some period of time before the TNx channel stabilizes. The process of spanning the analyzer consists of diluting bottled gas, using a calibrator and delivering to the analyzer. The operator then manually spans the TNx and NOx channels individually through the user interface. The following Tables outline the manual steps necessary to span the analyzer.

07271B DCN6646

17

Image 19
Contents For use with the T200 Operators Manual, P/N Manual AddendumSAN DIEGO, CA Page About this Manual This page intentionally left blank Safety Messages Mise EN Garde Consignes DE SécuritéTable of Contents 07271B DCN6646 Do not pressurize the Sample/Span/Zero port above ambient T201 Ammonia AnalyzerTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Outlined in the following equations Theory of OperationMinimizing PMT Drift Sample Filtration Special Considerations for Ammonia MeasurementT201 Operating Specifications T201 Analyzer SpecificationsThis page intentionally left blank Calibration Procedure Step No Action Comment Zeroing the AnalyzerSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank + + Configurable Analog Output OverviewStrip Chart HyperlinkAnalog Output Pin-Outs M501 NH3 Maintenance Maintenance ScheduleDescription Replacing the Catalytic CartridgeThermocouple Location Replacing the ThermocoupleTeledyne API Model T201 NH3 Analyzer Operator Manual Flow Check Autozero Flow CheckThis page intentionally left blank Alarm Alarms and CautionsAlarms and Cautions Tools ProcedureCalculating Moly Converter Efficiencies PurposeCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank To our Website or contact Sales for recent updates T201 Spare Parts ListPCA,TEC DRIVER,PMT,KB Assy & TEST, Spare PS38 Teledyne API Model T201 NH3 Analyzer Operator ManualThis page intentionally left blank