Teledyne API Model T201 NH3 Analyzer Operator Manual | Calibration Procedure |
| |||
| Table | ||||
|
|
|
|
|
|
| Step No. | Action | Comment |
|
|
|
|
|
| ||
| 1. | Press CALS | The T201 enters the calibrate mode from Sample | ||
|
|
| mode. The ammonia gas is supplied through the | ||
|
|
| SPAN inlet port on the rear panel of the M501 NH3. | ||
|
|
| Calculate the Efficiency factor as outlined in Steps 2 | ||
|
|
| in table 2.5. |
|
|
| 2 | Table 2.5 | Follow Steps 4 to 6 in table 2.5. |
|
|
|
|
|
| ||
Note | The response time of the analyzer increases when using the Zero/Span valve | ||||
|
| option. |
|
|
|
07271B DCN6646 | 23 |