Maintenance Schedule

4.3REPLACING THE THERMOCOUPLE

The following instructions provide the necessary information to replace the existing thermocouple. Figure 4-2is provided for reference.

You must obtain the following replacement parts kit from TAPI:

010820000 , Type K thermocouple

You will need the following tools:

Nutdriver, 5/16

Nutdriver, 11/32

Diagonal Cutter

Philips head Screwdriver #2

CAUTION

Avoid damage to the unit: use only Type K thermocouple in the

M501 NH3 converter. It can be distinguished by its wire colors: red and yellow. If you have any other thermocouple, do NOT install it; call TAPI to obtain the correct part.

Figure 4-2. Thermocouple Location

07271B DCN6646

29

Image 31
Contents For use with the T200 Operators Manual, P/N Manual AddendumSAN DIEGO, CA Page About this Manual This page intentionally left blank Safety Messages Mise EN Garde Consignes DE SécuritéTable of Contents 07271B DCN6646 Do not pressurize the Sample/Span/Zero port above ambient T201 Ammonia AnalyzerTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Outlined in the following equations Theory of OperationMinimizing PMT Drift Sample Filtration Special Considerations for Ammonia MeasurementT201 Operating Specifications T201 Analyzer SpecificationsThis page intentionally left blank Calibration Procedure Step No Action Comment Zeroing the AnalyzerSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank + + Configurable Analog Output OverviewStrip Chart HyperlinkAnalog Output Pin-Outs M501 NH3 Maintenance Maintenance ScheduleDescription Replacing the Catalytic CartridgeThermocouple Location Replacing the ThermocoupleTeledyne API Model T201 NH3 Analyzer Operator Manual Flow Check Autozero Flow CheckThis page intentionally left blank Alarm Alarms and CautionsAlarms and Cautions Tools ProcedureCalculating Moly Converter Efficiencies PurposeCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank To our Website or contact Sales for recent updates T201 Spare Parts ListPCA,TEC DRIVER,PMT,KB Assy & TEST, Spare PS38 Teledyne API Model T201 NH3 Analyzer Operator ManualThis page intentionally left blank