Calibration Procedure | Teledyne API Model T201 NH3 Analyzer Operator Manual |
Table
Important: Each channel (TNx and NOx) must be Spanned individually
Step No. | Action |
| Comment |
|
|
|
| ||
1. | Press CAL | The T201 enters the calibrate mode from sample | ||
|
| mode. Confirm that excess NO gas is flowing past the | ||
|
| sample port. |
| |
2. | Channel Selection | Press TNX or NOx then Press ENTR. | ||
3. | Range Selection | Press LOW or HIGH range then Press ENTR. | ||
|
| Always calibrate the LOW range, Section 3.0below. | ||
4. | Press CONC | If NOx was chosen previously then the following will | ||
|
| be displayed: |
| |
|
| NOx | NO CONV | Exit |
|
| Press NOx, If you are delivering 450 PPB of NO then | ||
|
| Enter 450 PPB, Press ENTR. |
| |
|
| Repeat the same steps for NO. |
| |
|
| Press EXIT |
| |
|
| Once stability has been achieved, |
| |
|
| Press SPAN, ENTR, EXIT |
| |
|
| NOTE: Both NOx and NO should read 450 PPB. | ||
|
| ____________________________________________ | ||
|
| If TNx was chosen previously then the following will | ||
|
| be displayed: |
| |
|
| TNx | CONV | Exit |
|
| Press TNx enter 450 PPB, Press ENTR | ||
|
| Press EXIT |
| |
|
| Once stability has been achieved, |
| |
|
| Press SPAN, ENTR, EXIT |
| |
|
| TNx should read 450 PPB. |
| |
|
|
| ||
5. | Check Slopes | The slopes should be close to 1.0 +/- 0.300 | ||
|
|
|
|
|