Calibration Procedure

Teledyne API Model T201 NH3 Analyzer Operator Manual

Table 2-3. NO Calibration Procedure – NO Gas through the SAMPLE Port

Important: Each channel (TNx and NOx) must be Spanned individually

Step No.

Action

 

Comment

 

 

 

 

1.

Press CAL

The T201 enters the calibrate mode from sample

 

 

mode. Confirm that excess NO gas is flowing past the

 

 

sample port.

 

2.

Channel Selection

Press TNX or NOx then Press ENTR.

3.

Range Selection

Press LOW or HIGH range then Press ENTR.

 

 

Always calibrate the LOW range, Section 3.0below.

4.

Press CONC

If NOx was chosen previously then the following will

 

 

be displayed:

 

 

 

NOx

NO CONV

Exit

 

 

Press NOx, If you are delivering 450 PPB of NO then

 

 

Enter 450 PPB, Press ENTR.

 

 

 

Repeat the same steps for NO.

 

 

 

Press EXIT

 

 

 

Once stability has been achieved,

 

 

 

Press SPAN, ENTR, EXIT

 

 

 

NOTE: Both NOx and NO should read 450 PPB.

 

 

____________________________________________

 

 

If TNx was chosen previously then the following will

 

 

be displayed:

 

 

 

TNx

CONV

Exit

 

 

Press TNx enter 450 PPB, Press ENTR

 

 

Press EXIT

 

 

 

Once stability has been achieved,

 

 

 

Press SPAN, ENTR, EXIT

 

 

 

TNx should read 450 PPB.

 

 

 

 

5.

Check Slopes

The slopes should be close to 1.0 +/- 0.300

 

 

 

 

 

18

07271B DCN6646

Image 20
Contents SAN DIEGO, CA Manual AddendumFor use with the T200 Operators Manual, P/N Page About this Manual This page intentionally left blank Safety Messages Consignes DE Sécurité Mise EN GardeTable of Contents 07271B DCN6646 T201 Ammonia Analyzer Do not pressurize the Sample/Span/Zero port above ambientTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Theory of Operation Outlined in the following equationsMinimizing PMT Drift Special Considerations for Ammonia Measurement Sample FiltrationT201 Analyzer Specifications T201 Operating SpecificationsThis page intentionally left blank Calibration Procedure Zeroing the Analyzer Step No Action CommentSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank Configurable Analog Output Overview Strip ChartHyperlink + +Analog Output Pin-Outs Maintenance Schedule M501 NH3 MaintenanceReplacing the Catalytic Cartridge DescriptionReplacing the Thermocouple Thermocouple LocationTeledyne API Model T201 NH3 Analyzer Operator Manual Autozero Flow Check Flow CheckThis page intentionally left blank Alarms and Cautions AlarmAlarms and Cautions Procedure Calculating Moly Converter EfficienciesPurpose ToolsCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank T201 Spare Parts List To our Website or contact Sales for recent updatesPCA,TEC DRIVER,PMT,KB Teledyne API Model T201 NH3 Analyzer Operator Manual Assy & TEST, Spare PS38This page intentionally left blank