Calibration Procedure

Teledyne API Model T201 NH3 Analyzer Operator Manual

The most important criterion an individual can avail themselves of when spanning an analyzer with ammonia gas is “patience”. Ammonia is a very sticky gas and the response of the analyzer/calibration system depends on a number of factors. The following lists some important points the user should be aware of when calibrating:

Is this the first time the analyzer/calibration system has seen ammonia gas in the last couple of days? If so, spanning to a fixed value will take considerably longer than normal: possibly 12 hours to completely stabilize.

Is the tubing that is delivering span gas from the calibration bottles, made of stainless steel? Is the tubing from the rear of the calibrator to the analyzer, made of stainless steel? Where possible use stainless steel tubing.

Note

The operator is responsible for delivering a set amount of ammonia to the

 

analyzer, calculating the converter efficiency of the M501 NH3 and entering

 

the efficiency value through the user interface, if required.

Is the environment where the analyzer resides undergoing large temperature swings (+/- 5 oC about the norm temperature)? If so the sample line can absorb and desorb ammonia at concentration rates that are distinguishable to the analyzer. If possible, heat the sample line to a constant maximum ambient temperature.

Does the sampling environment contain large amounts of particulates? If so, it may be necessary to add the sampling filter option. As outlined previously, this will affect the analyzer response.

One method to increase the response of both the sampling system and the analyzer is to flow a larger concentration of ammonia gas through the pneumatics for a couple of hours. For example, if the expected range is 500 PPB of ammonia, then flowing 1000 PPB will speed up the absorption which occurs throughout the pneumatics. The long delays in ammonia response occur when the analyzer hasn’t sampled the gas for some time. Ammonia desorbs from the pneumatics, these surfaces need to be rewetted before span stability is achieved.

Prior to conducting the ammonia calibration, confirm that the ammonia converter efficiency is initially set to 1.000 using the following menu key selections:

20

07271B DCN6646

Image 22
Contents For use with the T200 Operators Manual, P/N Manual AddendumSAN DIEGO, CA Page About this Manual This page intentionally left blank Safety Messages Consignes DE Sécurité Mise EN GardeTable of Contents 07271B DCN6646 T201 Ammonia Analyzer Do not pressurize the Sample/Span/Zero port above ambientTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Theory of Operation Outlined in the following equationsMinimizing PMT Drift Special Considerations for Ammonia Measurement Sample FiltrationT201 Analyzer Specifications T201 Operating SpecificationsThis page intentionally left blank Calibration Procedure Zeroing the Analyzer Step No Action CommentSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank Hyperlink Configurable Analog Output OverviewStrip Chart + +Analog Output Pin-Outs Maintenance Schedule M501 NH3 MaintenanceReplacing the Catalytic Cartridge DescriptionReplacing the Thermocouple Thermocouple LocationTeledyne API Model T201 NH3 Analyzer Operator Manual Autozero Flow Check Flow CheckThis page intentionally left blank Alarms and Cautions AlarmAlarms and Cautions Purpose ProcedureCalculating Moly Converter Efficiencies ToolsCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank T201 Spare Parts List To our Website or contact Sales for recent updatesPCA,TEC DRIVER,PMT,KB Teledyne API Model T201 NH3 Analyzer Operator Manual Assy & TEST, Spare PS38This page intentionally left blank