Chapter 4 Features and Functions

System-Related Operations

Self-Test

Apower-onself-test occurs automatically when you turn on the instrument. This limited test assures you that the instrument and all installed plug-in modules are operational. This self-test does not perform the extensive set of tests that are included as part of the complete self-test described below.

A complete self-test runs a series of tests and takes approximately

20 seconds to execute. If all tests pass, you can have a high confidence that the instrument and all installed plug-in modules are operational.

If the complete self-test is successful, PASS is displayed on the front panel. If the self-test fails, FAIL is displayed and the ERROR annunciator turns on. See the HP 34970A Service Guide for instructions on returning the instrument to Hewlett-Packard for service.

Front-PanelOperation: To perform the complete front-panel self-test,

4

hold down as you turn on the instrument and hold down the key until you hear a long beep. The self-test will begin when you release the key following the beep.

Remote Interface Operation:

*TST?

Returns “0” if the self-test is successful or “1” if it fails.