Chapter 4 Features and Functions

System-Related Operations

no

Aself-test occurs automatically when you turn on the instrument. This limited test assures you that the instrument and all installed plug-in modules are operational. This self-test does not perform the extensive set of tests that are included as part of the complete self-test described below.

A vx self-test runs a series of tests and takes approximately

20 seconds to execute. If all tests pass, you can have a high confidence that the instrument and all installed plug-in modules are operational.

If the complete self-test is successful, PASS is displayed on the front panel. If the self-test fails, FAIL is displayed and the ERROR annunciator

 

turns on. See the Uc :=EC7M fxvx T”wx

for instructions on

 

 

returning the instrument to Hewlett-Packard for service.

4

Sctxt”

: To perform the complete front-panel self-test,

 

hold down

as you turn on the instrument and “ w “x x

 

 

” “xt t z uxx

. The self-test will begin when you release

 

 

the key following the beep.

 

 

exx Vxytvx bxt”

:

 

 

*TST?

Returns “0” if the self-test is successful or “1” if it fails.

143

Page 142
Image 142
HP 34970A manual Turns on. See the Uc =EC7M fxvx Twx, Xt t z uxx, Exx Vxytvx bxt