Chapter 3 System Overview

Measurement Input

Totalizer The multifunction module has a 26-bit totalizer which can count pulses at a 100 kHz rate. You can manually read the totalizer count or you can configure a scan to read the count.

 

+IN

26 Bits

-IN

Channel 03

Totalize

 

Gate

 

Gate

You can configure the totalizer to count on the rising edge or falling edge of the input signal.

The maximum count is 67,108,863 (226- 1). The count rolls over to “0” after reaching the maximum allowed value.

You can configure the totalizer to read without affecting the count or reset the count to zero without losing any counts.