Chapter 4 Features and Functions

Scanning

xIn this configuration, the instrument sweeps through the scan list once each time a low-going TTL pulse is received

on the rear-panel R V

line (pin 6).

 

 

 

 

 

 

 

Ext Trig Input

Gnd

Ext Trig Connector

You can specify a scan count which sets the number of external

pulses the instrument will accept before terminating the scan.

4

See “Scan Count” on page 86 for more information.

 

If the instrument receives an external trigger before it is ready to accept one, it will buffer one trigger before generating an error.

All readings from the scan are stored in non-volatile memory. Readings accumulate in memory until the scan is terminated (until the scan count is reached or until you abort the scan).

Mx+B scaling and alarm limits are applied to measurements during the scan and all data is stored in non-volatile memory.

Front-Panel Operation:

EXTERNAL SCAN

To initiate the scan, press . The EXT annunciator turns on as a reminder that an External Scan is in progress. When a TTL pulse is received, the scan starts and readings are stored in memory. To stop

a scan, press and hold .

Remote Interface Operation: The following program segment configures the instrument for an External Scan.

TRIG:SOURCE EXT

Select the external trigger configuration

TRIG:COUNT 2

Sweep the scan list 2 times

INIT

Initiate the scan

NTo stop a scan, send the ABORt command.

83

Page 82
Image 82
HP 34970A Ext Trig Connector, See Scan Count on page 86 for more information, Select the external trigger configuration