Chapter 5 Remote Interface Reference

Status System Commands

Questionable Data Register Commands

gzz z wz v”z 9D7 “z z”z wz“

STATus:QUEStionable:CONDition?

Query the condition register in this register group. This is a read-only register and bits are not cleared when you read the register. A *RST (Factory Reset) will clear all bits in a condition register. A query of this register returns a decimal value which corresponds to the binary- weighted sum of all bits set in the register.

STATus:QUEStionable[:EVENt]?

Query the event register in this register group. This is a read-only register. Once a bit is set, it remains set until cleared by this command or *CLS (clear status) command. A query of this register returns a decimal value which corresponds to the binary-weighted sum of all bits set in the register.

STATus:QUEStionable:ENABle <zv>

STATus:QUEStionable:ENABle?

5

Enable bits in the enable register in this register group. The selected bits are then reported to the Status Byte. A *CLS (clear status) clear the enable register but it does clear all bits in the event register. A STATus:PRESet clears all bits in the enable register. To enable bits in the enable register, you must write a decimal value which corresponds to the binary-weighted sum of the bits you wish to enable in the register.

The :ENABle? query returns a decimal value which corresponds to the binary-weighted sum of all bits enabled by the STATus:QUES:ENABle command.

287

Page 285
Image 285
HP 34970A Status System Commands Questionable Data Register Commands, Gzz z wz vz 9D7 z zz wz, STATusQUEStionableEVENt?