Chapter 1 Quick Start

If the Instrument Does Not Turn On

1Remove the power cord. Remove the fuse-holder assembly from the rear panel.

2Remove the line-voltage selector from the assembly.

3Rotate the line-voltage selector until the correct voltage appears in the window.

Fuse: 500 mAT (for all line voltages)

HP Part Number: 2110-0458

4Replace the fuse-holder assembly in the rear panel.

100, 120 (127), 220 (230) or 240 Vac

Verify that the correct line voltage is selected and the power-line fuse is good.

28