Chapter 5 Remote Interface Reference

Frequency Configuration Commands

Frequency Configuration Commands

See also “Frequency Measurement Configuration” in chapter 4 starting on page 118.

[SENSe:]

FREQuency:RANGe:LOWer {320200MINMAX}[,(@<xt >)]

FREQuency:RANGe:LOWer? [{(@<x

>)MINMAX}]

Specify the lowest frequency expected in the input signal for frequency measurements on the specified channels. The instrument selects a slow, medium (default), or fast measurement timeout based on the frequency you specify. MIN = 3 Hz. MAX = 200 Hz. The :LOW? query returns the timeout setting on the specified channels. Returns “3”, “20”, or “200”.