The Plug-In Modules at a Glance

For complete specifications on each plug-in module, refer to the module sections in chapter 9.

HP 34901A 20-Channel Armature Multiplexer

20 channels of 300 V switching

Two channels for DC or AC current measurements (100 nA to 1A)

Built-in thermocouple reference junction

Switching speed of up to 60 channels per second

Connects to the internal multimeter

For detailed information and a module diagram, see page 164.

Each of the 20 channels switches both HI and LO inputs, thus providing fully isolated inputs to the internal multimeter. The module is divided into two banks of 10 two-wire channels each. When making four-wire resistance measurements, channels from Bank A are automatically paired with channels from Bank B. Two additional fused channels are included on the module (22 channels total) for making calibrated DC or AC current measurements with the internal multimeter (external shunt resistors are not required). You can close multiple channels on this module only if you have not configured any channels to be part of the scan list. Otherwise, all channels on the module are break-before-make.

HP 34902A 16-Channel Reed Multiplexer

16 channels of 300 V switching

Built-in thermocouple reference junction

Switching speed of up to 250 channels per second

Connects to the internal multimeter

For detailed information and a module diagram, see page 166.

Use this module for high-speed scanning and high-throughput automated test applications. Each of the 16 channels switches both HI and LO inputs, thus providing fully isolated inputs to the internal multimeter. The module is divided into two banks of eight two-wire channels each. When making four-wire resistance measurements, channels from Bank A are automatically paired with channels from Bank B. You can close multiple channels on this module only if you have not configured any channels to be part of the scan list. Otherwise, all channels on the module are break-before-make.

7