Scanning Overview

You can use either the internal DMM or an external DMM to make measurements of your configured channels. However, the instrument allows only one scan list at a time; you cannot scan some channels

using the internal DMM and others using an external DMM.

Readings are stored in HP 34970A memory when the internal DMM is used.

If the internal DMM is installed and enabled, the instrument will automatically use it for scanning. For externally-controlled scans, you must either remove the internal DMM from the HP 34970A or disable it (see “Internal DMM Disable” on page 145).

h}y

You can configure the event or action that controls the onset of each sweep through the scan list (a sweep is one pass through the scan list):

You can set the instrument’s internal timer to automatically scan at a specific interval. You can also program a time delay between channels in the scan list.

You can manually control a scan by repeatedly pressing from

 

the front panel.

5

You can start a scan by sending a software command from the

 

remote interface (MEASure? or INITiate command).

∙ You can start a scan when an external TTL trigger pulse is received.

∙ You can start a scan when an alarm event is logged on the channel being monitored.