Chapter 4 Features and Functions

System-Related Operations

mo

During a scan, the instrument stores all readings and alarms with the current time and date. The instrument stores the time and date information in non-volatile memory.

When shipped from the factory, the instrument is set to the current time and date (U.S. Mountain Time).

Sctxt”

:

TIME 03:45 PM

JUN 01 1997

exx Vxytvx bxt”: Use the following commands to set the time and date.

SYST:TIME

15,45,00

fxx _

4

 

SYST:DATE

1997,06,01

fxt Wx 83 8EEC

 

cg X

You can scan through the configured channels using either the internal DMM or an external instrument. For externally-controlled scans,

you must either remove the internal DMM from the instrument or disable it.

For information on controlling a scan with an external instrument, refer to “Scanning With External Instruments” on page 95.

When shipped from the factory, the internal DMM is enabled. When you change the state of the internal DMM, the instrument issues a Factory Reset (*RST command).

A Factory Reset (*RST command) or Instrument Preset (SYSTem:PRESet

command) wx affect the internal DMM configuration.