Chapter 4 Features and Functions

Alarm Limits

Remote Interface Operation (Digital Input Channel): To assign the alarm number to report any alarm conditions on the specified digital input channels, use the following command.

OUTPut:ALARm[1234]:SOURce (@<ch_list>)To configure alarms on the specified digital input channel, usethe following commands (also see the example on the following page).

CALCulate

:COMPare:TYPE {EQUalNEQual}[,(@<ch_list>)] :COMPare:DATA <data>[,(@<ch_list>)] :COMPare:MASK <mask>[,(@<ch_list>)]

Select EQUal to generate an alarm when the data read from the port is equal to CALC:COMP:DATA after being masked by CALC:COMP:MASK.

Select NEQual (not equal) to generate an alarm when the data read 4 from the port is not equal to CALC:COMP:DATA after being masked by CALC:COMP:MASK.

Use CALC:COMP:MASK to designate the “don’t care” bits. Bits that you set to “0” in the mask are ignored.

To enable the specified alarm mode, send the following command.CALCulate:COMPare:STATe ON [,(@<ch_list>)]