Chapter 9 Specifications

Measurement Rates and System Characteristics

u

Single Channel Measurement Rates [1] [2]

 

 

System Characteristics

Function

Resolution

Readings/s

 

Scan Triggering

 

DCV, 2-Wire Ohms:

612 (10 PLC)

6 (5)

 

Scan Count:

1 to 50,000 or continuous

 

512 (1 PLC)

57 (47)

 

Scan Interval:

0 to 99 hours; 1 ms step size

 

412 (0.02 PLC)

600

 

Channel Delay:

0 to 60 seconds/channel; 1 ms step size

 

0.1 °C (1 PLC)

57 (47)

 

External Trig Delay:

< 5 ms; With Monitor On, < 200 ms

Thermocouple:

 

External Trig Jitter:

< 2 ms

 

(0.02 PLC)

220

 

 

 

 

 

Alarms

 

 

 

 

 

 

RTD, Thermistor:

0.01 °C (10 PLC)

6 (5)

 

Alarm Outputs:

4 TTL compatible. Selectable TTL logic

 

 

HI or LO on Fail

 

0.1 °C (1 PLC)

57 (47)

 

 

 

 

Latency:

5 ms (typical)

 

1 °C (0.02 PLC)

220

 

 

 

 

 

ACV:

612 Slow (3 Hz)

0.14

 

Memory

Battery Backed, 4 year typical life [5]

 

Readings:

50,000 readings

 

612 Medium (20 Hz)

1

 

States:

5 instrument states

 

612 Fast (200 Hz)

8

 

Alarm Queue:

Up to 20 events

 

[3]

100

 

 

 

 

 

 

 

 

612

 

General Specifications

 

 

 

 

 

100 V / 120 V / 220 V / 240 V ±10%

Frequency, Period:

612 Digits (1s gate)

1

 

Power Supply:

 

Power Line Frequency:

45 Hz to 66 Hz automatically sensed

 

512 Digits (100 ms)

9

 

 

 

Power Consumption:

(12 W) 25 VA peak

 

412 Digits (10 ms)

70

 

 

 

Operating Environment:

Full accuracy for 0 °C to 55 °C

 

 

 

 

System Speeds [4]

 

 

 

 

Full accuracy to 80% R.H. at 40 °C

 

 

 

 

Storage Environment:

-40 °C to 70 °C [5]

 

 

 

INTO Memory

 

Ch/s

 

Weight (Mainframe):

Net: 3.6 kg (8.0 lbs)

Single Channel DCV

 

600

 

Safety:

Conforms to CSA, UL-1244, IEC 1010 Cat I

 

 

RFI and ESD:

CISPR 11, IEC 801/2/3/4

34902A Scanning DCV

 

250

 

 

 

Warranty:

3 years

34907A Scanning Digital Input

250

 

 

 

 

34902A Scanning DCV, scaling and 1 alarm fail

220

 

 

 

34907A Scanning Totalize

170

 

 

 

34902A Scanning Temperature

160

 

[1] Reading speeds for 60 Hz and (50 Hz) operation; autozero OFF

34902A Scanning ACV [3]

100

 

[2] For fixed function and range, readings to memory,

34902A Scanning DCV/Ohms, alternate channels

90

 

scaling and alarms off, autozero OFF

34901A/34908A Scanning DCV

60

 

[3] Maximum limit with default settling delays defeated

INTO and OUT of Memory to HP-IB or RS-232 (INIT, FETCh)

 

[4] Speeds are for 412 digits, delay 0, display off, autozero off.

 

Using 115 kbaud RS-232 setting.

34902A Scanning DCV

 

180

 

 

 

 

°

34902A Scanning DCV with Time stamp

150

 

[5] Storage at temperatures above 40 C will decrease battery life

 

 

 

OUT of Memory to HP-IB

 

 

 

 

 

Readings

 

800

 

 

 

Readings with Time stamp

450

 

 

 

Readings with all Format Options ON

310

 

 

 

OUT of Memory to RS-232

 

 

 

 

Readings

 

600

 

 

 

Readings with Time stamp

320

 

 

 

Readings with all Format Options ON

230

 

 

 

DIRECT to HP-IB or RS-232

 

 

 

 

Single Channel DCV

 

440

 

 

 

34902A Scanning DCV

 

200

 

 

 

Single Channel MEAS DCV 10 or MEAS DCV 1

25

 

 

 

Single Channel MEAS DCV or MEAS OHMS

12

 

 

 

 

 

 

 

 

 

IDP

Page 403
Image 403
HP 34970A manual Measurement Rates and System Characteristics, Single Channel Measurement Rates 1 System Characteristics