Chapter 5 Remote Interface Reference

Totalizer Commands

Totalizer Commands

See also “Totalizer Operations” in chapter 4 starting on page 135.

MEASure:TOTalize? {READRRESet} ,(@<scan_list>)

Configure the instrument to read the count on the specified totalizer channels on the multifunction module and immediately sweep through the scan list one time. Note that this command also redefines the scan list. The readings are sent directly to the instrument’s output buffer but the readings are not stored in reading memory. The totalizer channel is numbered s03”, where s represents the slot number.

To read the totalizer during the scan without resetting the count, select the READ parameter. To read the totalizer during the scan and reset the count to “0” after it is read, select the RRESet parameter (this means “read and reset”).

CONFigure:TOTalize {READRRESet} ,(@<scan_list>)

Configure the instrument to read the specified totalizer channels on the multifunction module but do not initiate the scan. Note that this command also redefines the scan list. The totalizer channel is numbered s03”, where s represents the slot number.

To read the totalizer during the scan without resetting the count, select the READ parameter. To read the totalizer during the scan and reset the count to “0” after it is read, select the RRESet parameter (this means “read and reset”).

[SENSe:]

TOTalize:TYPE {READRRESet}[,(@<ch_list>)] TOTalize:TYPE? [(@<ch_list>)]

Disable (default) or enable an automatic reset of the count when the specified totalizer channels are scanned. To read the count without resetting the value, select the READ parameter. To read the totalize and reset the count to “0” after it is read, select the RRESet parameter (this means “read and reset”). The totalizer channel is numbered s03”, where s represents the slot number. The :TYPE? query returns the reset configuration currently selected. Returns “READ” or “RRES”.