Chapter 4 Features and Functions

Scanning

U

If you do not specify a channel delay, the instrument selects a delay for you. The delay is determined by function, range, integration time, and ac filter setting as shown below.

DC Voltage, Thermocouple, DC Current (for all ranges):

Integration Time

Channel Delay

PLC > 1

2.0 ms

PLC 1

1.0 ms

Resistance, RTD, Thermistor (2- and 4-wire):

Range

Channel Delay

(For PLC > 1)

 

100Ω

2.0 ms

1 kΩ

2.0 ms

10 kΩ

2.0 ms

100 kΩ

25 ms

1 MΩ

30 ms

10 MΩ

200 ms

100 MΩ

200 ms

Range

Channel Delay

(For PLC 1)

 

100Ω

1.0 ms

1 kΩ

1.0 ms

10 kΩ

1.0 ms

100 kΩ

20 ms

1 MΩ

25 ms

10 MΩ

200 ms

100 MΩ

200 ms

4

AC Voltage, AC Current (for all ranges):

AC Filter

Channel Delay

Slow (3 Hz)

7.0 sec

Medium (20 Hz)

1.0 sec

Fast (200 Hz)

120 ms

Frequency, Period:

AC Filter

Channel Delay

Slow (3 Hz)

0.6 sec

Medium (20 Hz)

0.3 sec

Fast (200 Hz)

0.1 sec

Digital Input, Totalize:

Channel Delay

0 sec

89

Page 88
Image 88
HP 34970A manual Integration Time Channel Delay, Range Channel Delay, AC Filter Channel Delay