Chapter 4 Features and Functions

Frequency Measurement Configuration

Frequency Measurement Configuration

To connect an ac source to the module’s screw terminals, see page 21.

f ]

The instrument uses three different timeout ranges for frequency measurements. The instrument selects a slow, medium, or fast timeout based on the input frequency that you specify for the selected channels.

Input Frequency

Timeout

 

 

3 Hz to 300 kHz (Slow)

1 s

20 Hz to 300 kHz (Medium)

100 ms

200 Hz to 300 kHz (Fast)

10 ms

 

 

The instrument selects the medium timeout (20 Hz) when the function is changed or after a Factory Reset (*RST command).

An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.

Front-Panel Operation: First, select the frequency function on the active channel. Then, go to the Advanced menu and select the slow timeout (3 Hz), medium timeout (CA b ), or fast timeout (200 Hz) for the active channel. The default is the medium timeout value.

LF 3 HZ:SLOW

Remote Interface Operation: Specify the lowest frequency expected in the input signal on the specified channels. The instrument selects the appropriate timeout based on the frequency you specify (see table above). The MEASure? and CONFigure commands automatically select the 20 Hz (medium) timeout.

SENS:FREQ:RANG:LOW 3,(@203) Select the slow timeout (3 Hz)

118

Page 117
Image 117
HP 34970A manual Frequency Measurement Configuration, Sensfreqranglow 3,@203 Select the slow timeout 3 Hz