Chapter 4 Features and Functions

Temperature Measurement Configuration

o g

To connect a thermistor to the module’s screw terminals, see page 21.

The instrument supports 2.2 kΩ (44004), 5 kΩ (44007), and 10 kΩ (44006) thermistors.

Front-Panel Operation: To select the thermistor function for the active channel, choose the following items.

TEMPERATURE , THERMISTOR

To select the thermistor type for the active channel, choose from the following items.

TYPE 2.2 KOHM , TYPE 5 KOHM , TYPE 10 KOHM

Remote Interface Operation: You can use the MEASure? or

CONFigure command to select the probe type and thermistor type.

For example, the following statement configures channel 301 for measurements of a 5 kΩ thermistor:

CONF:TEMP THER,5000,(@301)

You can also use the SENSe command to select the probe type and thermistor type. For example, the following statement configures channel 103 for measurements of a 10 kΩ thermistor:

SENS:TEMP:TRAN:THERM:TYPE 10000,(@103)

112

Page 111
Image 111
HP 34970A manual To connect a thermistor to the module’s screw terminals, see