Chapter 4 Features and Functions

Totalizer Operations

exx Vxytvx bxt”: To read the count from the specified totalizer channel, send the following command. The count may be returned with time stamp, channel number, and alarm status information depending on the FORMat:READing command setting (see “Reading Format” on page 87 for more information).

SENS:TOT:DATA? (@303)

To configure the totalizer reset mode, send either of the following commands (RRESet means “read and reset”).

SENSe:TOTalize:TYPE {READRRESet}[,(@<v“r

>)]

 

CONFigure:TOTalize {READRRESet} ,(@<r

>)

 

To configure the totalizer to count on the falling edge (zt”x

) or

 

rising edge (

) of the input signal, send the following command.

4

 

 

 

SENSe:TOTalize:SLOPe {NEGPOS} ,[(@<v“r >)]

To immediately clear the count on the specified totalizer channel (whether scanning or not), send the following command.

SENSe:TOTalize:CLEar:IMMediate [(@<v“r >)]

137

Page 136
Image 136
HP 34970A manual 137