Chapter 9 Specifications

AC Measurement and Operating Characteristics

gi u

AC Measurement Characteristics [1]

True RMS AC Voltage

AC-coupled True RMS – measures

Measurement Method:

 

the ac component of input with up

 

to 300 Vdc of bias on any range

Crest Factor:

Maximum 5:1 at Full Scale

Additional Crest Factor

 

 

Errors (non-sinewave): [2]

Crest Factor 1-2: 0.05% of reading

 

Crest Factor 2-3: 0.15% of reading

 

Crest Factor 3-4: 0.30% of reading

AC Filter Bandwidth:

Crest Factor 4-5: 0.40% of reading

3 Hz 300 kHz

Slow

Medium

20 Hz

300 kHz

Fast

200 Hz

300 kHz

Input Impedance:

1 MΩ ± 2%, in parallel with 150 pF

Input Protection:

300 Vrms on all ranges

 

 

 

Frequency and Period

Reciprocal counting technique

Measurement Method:

Voltage Ranges:

Same as AC Voltage function

Gate Time:

1s, 100 ms, or 10 ms

Measurement Timeout:

Selectable 3 Hz, 20 Hz, 200 Hz LF limit

 

 

 

True RMS AC Current

 

 

Measurement Method:

Direct coupled to the fuse and

 

shunt. AC-coupled True RMS

 

measurement (measures the

 

ac component only)

Shunt Resistance:

5Ω for 10 mA; 0.1Ω for 100 mA, 1A

Input Protection:

1.5A 250 V fuse on HP 34901A module

 

 

Measurement Noise Rejection [3]

 

AC CMRR:

70 dB

 

 

 

 

Measurement Considerations (Frequency and Period)

All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors.

AC Operating Characteristics [4]

Function

Digits [5]

Readings/s

AC Filter

ACV, ACI:

612

7 sec/reading

Slow (3 Hz)

 

612

1

Medium (20 Hz)

 

612

8 [6]

Fast (200 Hz)

 

612

10

Fast (200 Hz)

 

612

100 [7]

Fast (200 Hz)

Single Channel Measurement Rates [8]

 

Function

Resolution

Readings/s

ACV:

612

Slow (3 Hz)

0.14

 

61

2

Medium (20 Hz)

1

 

61

2

Fast (200 Hz)

8

 

61

2

[7]

100

 

 

Frequency, Period:

612

Digits (1s gate)

0.77

 

61

2

Digits (1s gate) [7]

1

 

51

2

Digits (100 ms)

2.5

 

51

2

Digits (100 ms) [7]

9

 

41

2

Digits (10 ms)

3.2

 

41

2

Digits (10 ms) [7]

70

[1]300 Vrms isolation voltage (ch-ch, ch-earth)

[2]For frequencies below 100 Hz, slow AC filter specified for sinewave input only

[3]For 1 kΩ unbalance in LO lead

[4]Maximum reading rates for 0.01% of ac step additional error. Additional settling delay required when input dc level varies.

[5]612 digits=22 bits, 512 digits=18 bits, 412 digits=15 bits

[6]For external trigger or remote operation using default settling delay (Delay Auto)

[7]Maximum limit with default settling delays defeated

[8]For fixed function and range, readings to memory, scaling and alarms turned off

IDO 9

Page 402
Image 402
HP 34970A manual Gi u, AC Measurement and Operating Characteristics