Chapter 6 Error Messages

Instrument Errors

501

I/O processor: isolator framing error

502

I/O processor: isolator overrun error

511

Communications: RS-232 framing error

512

Communications: RS-232 overrun error

513

Communications: RS-232 parity error

514

RS-232 only: unable to execute using HP-IB

 

There are three commands which are allowed only with the RS-232

 

interface: SYSTem:LOCal, SYSTem:REMote, and SYSTem:RWLock.

521

Communications: input buffer overflow

522

Communications: output buffer overflow

532

Not able to achieve requested resolution

 

The instrument cannot achieve the requested measurement resolution.

 

You may have specified an invalid resolution in the CONFigure or

 

MEASure? command.

540

Not able to null channel in overload

 

The instrument cannot store an overload reading (9.90000000E+37)

 

as the offset for Mx+B scaling using a null measurement.

550

Not able to execute command in local mode

 

The instrument has received a READ? or MEASure? command while in

 

the local mode.