Chapter 9 Specifications

Interpreting Internal DMM Specifications

bdR u

o 9=4 ””” ”Vhh’s relative accuracy over its full measurement range for short time intervals and within a stable environment. Short-term accuracy is usually specified for a 24-hour period and for a ±1 °C temperature range.

lYR u

These long-term accuracy specifications are valid for a 23 °C ± 5 °C temperature range. These specifications include the initial calibration errors plus the internal Vhh’s long-term drift errors.

y

Accuracy is usually specified for a 23 °C ± 5 °C temperature range. This is a common temperature range for many operating environments. You must add additional temperature coefficient errors to the accuracy specification if you are operating the internal Vhh outside a 23 °C ± 5 °C temperature range (the specification is per °C).

418