Chapter 5 Remote Interface Reference

The MEASure? and CONFigure Commands

MEASure:DIGital:BYTE? (@< >)

Configure the instrument to read the specified digital input channels on the multifunction module and immediately sweep through the scan list

one time. b v xvy vz. The readings

are sent directly to the instrument’s output buffer but the readings vz stored in reading memory. The digital input channels are numbered

“ 01” (LSB) and “ 02” (MSB), where represents the slot number.

Note that if you include both digital input channels in the scan list, the instrument will read data from both ports simultaneously with the same time stamp. This will allow you to externally combine the two 8-bit value into one 16-bit value.

MEASure:TOTalize? {READRRESet} ,(@< >)

Configure the instrument to read the count on the specified totalizer channels on the multifunction module and immediately sweep through

the scan list one time. b v xvy vz. The readings are sent directly to the instrument’s output buffer but the

readings vz

stored in reading memory. The totalizer channel is

numbered

03”, where represents the slot number.

To read the totalizer during the scan without resetting the count, select the READ parameter. To read the totalizer during the scan and reset the count to “0” after it is read, select the RRESet parameter (this means “read and reset”).

210