Chapter 5 Remote Interface Reference

Simplified Programming Overview

Example: Using MEASure?

The following program segment shows how to use the MEASure? command to make a measurement on one channel. This example configures the instrument for dc voltage measurements, internally triggers the instrument to scan one channel, and then sends the reading to the instrument’s output buffer.

MEAS:VOLT:DC? 10,0.003,(@301)

This is the simplest way to take a reading. However, you do not have any flexibility with MEASure? to set the scan count, channel delay, etc. All measurement parameters except function, range, and resolution are preset for you automatically (see the table on page 201).

Example: Using CONFigure With READ?

The following program segment shows how to use the READ? command with CONFigure to make an externally-triggered scan on one channel. The program configures the instrument for a dc voltage measurement. Using CONFigure does not place the instrument in the “wait-for-

trigger” state. The READ? command places the instrument in the5 “wait-for-trigger” state, scans the channel once when the Ext Trig

terminal is pulsed on the rear panel, and sends the reading to the instrument’s output buffer.

CONF:VOLT:DC 10,0.003,(@301)

TRIG:SOUR EXT

READ?