Chapter 5 Remote Interface Reference

Resistance Configuration Commands

Resistance Configuration Commands

See also “Resistance Measurement Configuration” in chapter 4 starting on page 115.

[SENSe:]

RESistance:OCOMpensated {OFFON}[,(@<ch_list>)] RESistance:OCOMpensated? [(@<ch_ list>)] FRESistance:OCOMpensated {OFFON}[,(@<ch_list>)] FRESistance:OCOMpensated? [(@<ch _list>)]

Disable or enable offset-compensation for resistance measurements. Once enabled, offset compensation is applied to both 2-wire and 4-wire resistance measurements. The default is “OFF”. The :OCOM? query returns the offset compensation setting on the specified channels. Returns “0” (OFF) or “1” (ON).

Current Configuration Commands

See also “Current Measurement Configuration” in chapter 4 starting on page 116.

Note: Current measurements are allowed only on channels 21 and 22 on the HP 34901A multiplexer module.

[SENSe:]

CURRent:AC:BANDwidth {320200MINMAX}[,(@<ch_list>)] CURRent:AC:BANDwidth? [{(@<ch_list>)MINMAX}]

Specify the lowest frequency expected in the input signal for ac current measurements on the specified channels. The instrument selects the slow, medium (default), or fast ac filter based on the frequency you specify. MIN = 3 Hz. MAX = 200 Hz. The :BAND? query returns the ac filter setting on the specified channels. Returns “3”, “20”, or “200”.

224