Chapter 5 Remote Interface Reference

The MEASure? and CONFigure Commands

CONFigure:VOLTage:DCCONFigure:VOLTage:AC

[{<range>AUTOMINMAXDEF}

[,<resolution>MINMAXDEF}],] (@<scan_list>)

Configure the specified channels for dc or ac voltage measurements but do not initiate the scan. Note that this command also redefines the scan list. For ac measurements, the resolution is actually fixed at 612 digits;

the resolution parameter only affects the number of digits shown on the front-panel.

CONFigure:RESistanceCONFigure:FRESistance

[{<range>AUTOMINMAXDEF}

[,<resolution>MINMAXDEF}],] (@<scan_list>)

Configure the specified channels for 2-wire or 4-wire measurements but do not initiate the scan. Note that this command also redefines the scan list.

For 4-wire measurements (FRES), the instrument automatically pairs channel n with channel n+10 (34901A) or n+8 (34902A) to provide the source and sense connections. For example, make the source connections to the HI and LO terminals on channel 2 and the sense connections to the HI and LO terminals on channel 12. Specify the paired channel in the lower bank (source) as the scan_list channel.

CONFigure:CURRent:DCCONFigure:CURRent:AC

[{<range>AUTOMINMAXDEF}

[,<resolution>MINMAXDEF}],] (@<scan_list>)

Note: Current measurements are allowed only on channels 21 and 22 on the HP 34901A multiplexer module.

Configure the specified channels for dc or ac current measurements but do not initiate the scan. Note that this command also redefines the scan list. For ac measurements, the resolution is actually fixed at 612 digits;

the resolution parameter only affects the number of digits shown on the front-panel.

212