Chapter 4 Features and Functions

Alarm Limits

For details on configuring alarms on the multifunction module, see “Using Alarms With the Multifunction Module” on page 130.

A Factory Reset (*RST command) clears all alarm limits and turns off all alarms. An Instrument Preset (SYSTem:PRESet command) or

Card Reset (SYSTem:CPON command) wx

clear the alarm limits

and wx

turn off alarms.

 

Sctxt”: To select the alarm for use on the active channel, choose from the following items.

NO ALARM , USE ALARM 1 , ... USE ALARM 4

Then, choose from the following alarm conditions.

HI+LO ALARMS , HI ALARM ONLY , LO ALARM ONLY

4

Then, set the desired limit values and exit the menu. Note that the

instrument does not start evaluating the alarm conditions until you

 

exit the Mt

menu.

 

exx Vxytvx bxt”: To assign the alarm number to report any alarm conditions on the specified channels, use the following command (if not assigned, all alarms on all channels are reported on Alarm 1 by default).

OUTPUT:ALARM2:SOURCE (@103,212)

To set the upper and lower alarm limits on the specified channels, use the following commands.

CALC:LIMIT:UPPER 5.25,(@103,212)

CALC:LIMIT:LOWER 0.025,(@103,212)

To enable the upper and lower alarm limits on the specified channels, use the following commands.

CALC:LIMIT:UPPER:STATE ON,(@103,212)

CALC:LIMIT:LOWER:STATE ON,(@103,212)