Chapter 4 Features and Functions

General Measurement Configuration

The specified resolution is used for all measurements on the selected channel. If you have applied Mx+B scaling or have assigned alarms to the selected channel, those measurements are also made using the specified resolution. Measurements taken during the Monitor function also use the specified resolution.

Changing the number of digits does more than just change the resolution of the instrument. It also changes the ”t”x3

which is the period the instrument’s analog-to-digital (A/D) converter samples the input signal for a measurement. See “Custom A/D Integration Time,” on page 103 for more information.

The MEASure? and CONFigure commands contain an optional resolution parameter which allows you to specify the resolution.

The instrument returns to 512 digits when the measurement function

is changed and after a Factory Reset (*RST command). An Instrument

Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON 4 command) does not change the resolution setting.

Front-Panel Operation: First, select the measurement function on the active channel. You are automatically guided to the next level of

the menu where you can select the number of digits. The default is 512 digits.

6 1/2 DIGITS

For temperature measurements, go to the menu and select the number of digits displayed past the decimal point on the selected channel.

DISPLAY 1 °C

101

Page 100
Image 100
HP 34970A manual 101