Chapter 5 Remote Interface Reference

The SCPI Status System

ly QQm2

You can use the Status Byte “Message Available” bit (bit 4) to determine when data is available to read into your computer. The instrument subsequently clears bit 4 only after all messages have been read from the output buffer.

i Sf

1.Send a Device Clear message to clear the instrument’s output buffer (e.g., CLEAR 709).

2.Clear the event registers using the *CLS command.

3.Set up the enable register masks. Execute the *ESE command to configure the Standard Event enable register and the *SRE command to configure the Status Byte enable register.

4.Send the *OPC? command and enter the result to ensure synchronization.

5.Enable your computer’s IEEE-488 SRQ interrupt.

i y S S

5

1. Send a Device Clear message to clear the instrument’s output buffer (e.g., CLEAR 709).

2. Clear the event registers using the *CLS command.

3. Enable the “Operation Complete” bit (bit 0) in the Standard Event register using the *ESE 1 command.

4. Send the *OPC? command and enter the result to ensure synchronization.

5. Execute your command string to program the desired configuration, and then send the *OPC command as the last command (note that if a scan is in progress, the *OPC command will wait until the zz

scan is complete). When the command sequence is completed, the “Operation Complete” bit (bit 0) is set in the Standard Event register.

6. Use a Serial Poll to check to see when bit 5 (routed from the Standard Event register) is set in the Status Byte condition register. You could also configure the instrument for an SRQ interrupt by sending *SRE 32 (Status Byte enable register, bit 5).