Chapter 9 Specifications

AC Measurement and Operating Characteristics

AC Measurement and Operating Characteristics

AC Measurement Characteristics [1]

True RMS AC Voltage

AC-coupled True RMS – measures

Measurement Method:

 

the ac component of input with up

 

to 300 Vdc of bias on any range

Crest Factor:

Maximum 5:1 at Full Scale

Additional Crest Factor

 

Errors (non-sinewave): [2]

Crest Factor 1-2: 0.05% of reading

 

Crest Factor 2-3: 0.15% of reading

 

Crest Factor 3-4: 0.30% of reading

 

Crest Factor 4-5: 0.40% of reading

AC Filter Bandwidth:

3 Hz 300 kHz

Slow

Medium

20 Hz 300 kHz

Fast

200 Hz 300 kHz

Input Impedance:

1 MΩ ± 2%, in parallel with 150 pF

Input Protection:

300 Vrms on all ranges

 

 

Frequency and Period

Reciprocal counting technique

Measurement Method:

Voltage Ranges:

Same as AC Voltage function

Gate Time:

1s, 100 ms, or 10 ms

Measurement Timeout:

Selectable 3 Hz, 20 Hz, 200 Hz LF limit

 

 

True RMS AC Current

Direct coupled to the fuse and

Measurement Method:

 

shunt. AC-coupled True RMS

 

measurement (measures the

 

ac component only)

Shunt Resistance:

5for 10 mA; 0.1for 100 mA, 1A

Input Protection:

1.5A 250 V fuse on 34901A module

 

Measurement Noise Rejection [3]

AC CMRR:

70 dB

 

 

Measurement Considerations (Frequency and Period)

All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors.

AC Operating Characteristics [4]

Function

Digits [5]

Readings/s

AC Filter

ACV, ACI:

612

 

7 sec/reading

Slow (3 Hz)

 

612

 

 

1

Medium (20 Hz)

 

612

 

 

8 [6]

Fast (200 Hz)

 

612

 

 

10

Fast (200 Hz)

612üxç 31ç 33

Çä

32 é

100 [7]

Fast (200 Hz)

 

 

Single Channel Measurement Rates [8]

 

Function

Resolution

 

Readings/s

ACV:

612

Slow (3 Hz)

0.14

 

61

2

Medium (20 Hz)

1

 

61

2

Fast (200 Hz)

8

 

61

2üxç 31ç 33 Çä

32 é [7]

100

Frequency, Period:

612

Digits (1s gate)

0.77

 

61

2

Digits (1s gate) [7]

1

 

51

2

Digits (100 ms)

2.5

 

51

2

Digits (100 ms) [7]

9

 

41

2

Digits (10 ms)

3.2

 

41

2

Digits (10 ms) [7]

70

[1]300 Vrms isolation voltage (ch-ch, ch-earth)

[2]For frequencies below 100 Hz, slow AC filter specified for sinewave input only

[3]For 1 kunbalance in LO lead

[4]Maximum reading rates for 0.01% of ac step additional error. Additional settling delay required when input dc level varies.

[5]612 digits=22 bits, 512 digits=18 bits, 412 digits=15 bits

[6]For external trigger or remote operation using default settling delay (Delay Auto)

[7]Maximum limit with default settling delays defeated

[8]For fixed function and range, readings to memory, scaling and alarms turned off

4079

Page 409
Image 409
Agilent Technologies 34970A manual AC Measurement and Operating Characteristics

34970A specifications

Agilent Technologies 34970A is a versatile data acquisition and measurement system that has gained popularity in various industries, including research and development, manufacturing, and education. The primary purpose of the 34970A is to offer an efficient solution for data collection, monitoring, and analysis, which can significantly enhance productivity and accuracy in testing applications.

One of the main features of the 34970A is its modular design, allowing users to customize the system according to their specific measurement needs. The mainframe can accommodate up to three plug-in measurement modules, which can include various types of measurements such as analog, digital, temperature, and frequency. This modularity provides flexibility for users to tailor the system to their requirements, making it highly adaptable to different applications.

Another notable characteristic of the Agilent 34970A is its impressive channel configuration. The system can support up to 120 measurement channels when fully configured, enabling extensive data acquisition without the need for multiple devices. This capability is essential for applications requiring simultaneous monitoring of multiple parameters or locations.

The 34970A utilizes advanced signal processing technologies to ensure high-accuracy measurements. The internal 6.5-digit resolution multimeter provides precise readings, while the device also supports various input types, such as thermocouples and RTDs for temperature measurements. The ability to perform mathematical functions, such as summation and averaging, on the acquired data further enhances its usability.

For data storage and management, the Agilent 34970A features built-in memory for storing up to 20,000 readings. Users can also easily transfer data to a PC through the RS-232 interface or GPIB (IEEE 488) for further analysis and reporting. Additionally, it offers the capability to program automatic data logging and scheduling of measurements, which streamlines the testing process and reduces manual intervention.

In terms of user interface, the 34970A is equipped with a large graphical display that provides clear visualization of measurement data and easy navigation through settings and options. The menu-driven interface makes it accessible for users of all skill levels, simplifying the setup and operation of the device.

Overall, the Agilent Technologies 34970A stands out for its modularity, high channel capacity, advanced measurement technologies, and user-friendly interface. As a reliable and efficient data acquisition system, it is well-suited for a wide range of applications, making it an essential tool for engineers and technicians looking to enhance measurement precision and efficiency.