Chapter 9 Specifications

Module Specifications

Module Specifications

34901A, 34902A, 34908A, 34903A, 34904A

 

 

 

 

Multiplexer

 

Actuator

Matrix

 

 

 

 

 

 

 

 

General

 

 

34901A

34902A

34908A

34903A

34904A

Number of Channels

 

20+2

16

40

20

4x8

 

 

 

 

 

 

 

 

 

 

 

2/4 wire

2/4 wire

1 wire

SPDT

2 wire

 

 

 

 

 

 

 

Connects to Internal DMM

 

Yes

Yes

Yes

No

No

Scanning Speed [1]

 

60 ch/s

250 ch/s

60 ch/s

 

 

Open/Close Speed

 

120/s

120/s

70/s

120/s

120/s

 

 

 

 

 

 

 

 

Maximum Input

 

 

 

 

 

 

 

Voltage (dc, ac rms)

 

300 V

300 V

300 V

300 V

300 V

Current (dc, ac rms)

 

1 A

50 mA

1 A

1 A

1 A

 

 

 

 

 

 

 

 

Power (W, VA)

 

 

50 W

2 W

50 W

50 W

50 W

 

 

 

 

 

 

 

Isolation (ch-ch, ch-earth)

dc, ac rms

300 V

300 V

300 V

300 V

300 V

DC Characteristics

 

 

 

 

 

 

 

Offset Voltage [2]

 

 

< 3 V

< 6 V

< 3 V

< 3 V

< 3 V

Initial Closed Channel R [2]

 

< 1

< 1

< 1

< 0.2

< 1

Isolation (ch-ch, ch-earth)

 

> 10 G

> 10 G

> 10 G

> 10 G

> 10 G

AC Characteristics

 

 

 

 

 

 

 

Bandwidth

 

 

10 MHz

10 MHz

10 MHz

10 MHz

10 MHz

 

 

 

 

 

 

 

Ch-Ch Cross Talk (dB) [3]

10 MHz

-45

-45

-18 [4]

-45

-33

Capacitance

 

HI to LO

< 50 pF

< 50 pF

< 50 pF

< 10 pF

< 50 pF

 

 

 

 

 

 

 

 

Capacitance

 

LO to Earth

< 80 pF

< 80 pF

< 80 pF

< 80 pF

< 80 pF

 

 

 

 

 

 

 

 

Volt-Hertz Limit

 

 

108

108

108

108

108

Other

 

 

 

 

 

 

 

T/C Cold Junction Accuracy [2] [5] (typical)

0.8 °C

0.8 °C

0.8 °C [7]

 

 

Switch Life

No Load (typical)

100M

100M

100M

100M

100M

 

 

 

 

 

 

 

Switch Life

Rated Load (typical) [6]

100k

100k

100k

100k

100k

Temperature

 

Operating

 

All Modules – 0 °C to 55 °C

 

 

 

 

 

 

 

Temperature

 

Storage

 

All Modules – -20 °C to 70 °C

 

Humidity (non-condensing)

 

 

All Modules – 40 °C / 80% R.H.

 

 

 

 

 

 

 

 

 

[1]Speeds are for 412 digits, delay 0, display off, autozero off. Using 115 kbaud RS-232 setting.

[2]Errors included in the DMM measurement accuracy specifications

[3]50source, 50load

[4]Isolation within channel 1 to 20 or 21 to 40 banks is -40 dB

[5]Thermocouple specifications not guaranteed when 34907A module is present

[6]Applies to resistive loads only

[7]Thermocouple measurements not recommended with 34908A module due to common LO configuration.

4099

Page 411
Image 411
Agilent Technologies 34970A manual Module Specifications

34970A specifications

Agilent Technologies 34970A is a versatile data acquisition and measurement system that has gained popularity in various industries, including research and development, manufacturing, and education. The primary purpose of the 34970A is to offer an efficient solution for data collection, monitoring, and analysis, which can significantly enhance productivity and accuracy in testing applications.

One of the main features of the 34970A is its modular design, allowing users to customize the system according to their specific measurement needs. The mainframe can accommodate up to three plug-in measurement modules, which can include various types of measurements such as analog, digital, temperature, and frequency. This modularity provides flexibility for users to tailor the system to their requirements, making it highly adaptable to different applications.

Another notable characteristic of the Agilent 34970A is its impressive channel configuration. The system can support up to 120 measurement channels when fully configured, enabling extensive data acquisition without the need for multiple devices. This capability is essential for applications requiring simultaneous monitoring of multiple parameters or locations.

The 34970A utilizes advanced signal processing technologies to ensure high-accuracy measurements. The internal 6.5-digit resolution multimeter provides precise readings, while the device also supports various input types, such as thermocouples and RTDs for temperature measurements. The ability to perform mathematical functions, such as summation and averaging, on the acquired data further enhances its usability.

For data storage and management, the Agilent 34970A features built-in memory for storing up to 20,000 readings. Users can also easily transfer data to a PC through the RS-232 interface or GPIB (IEEE 488) for further analysis and reporting. Additionally, it offers the capability to program automatic data logging and scheduling of measurements, which streamlines the testing process and reduces manual intervention.

In terms of user interface, the 34970A is equipped with a large graphical display that provides clear visualization of measurement data and easy navigation through settings and options. The menu-driven interface makes it accessible for users of all skill levels, simplifying the setup and operation of the device.

Overall, the Agilent Technologies 34970A stands out for its modularity, high channel capacity, advanced measurement technologies, and user-friendly interface. As a reliable and efficient data acquisition system, it is well-suited for a wide range of applications, making it an essential tool for engineers and technicians looking to enhance measurement precision and efficiency.