Chapter 6 Error Messages

Instrument Errors

 

Instrument Errors

 

111

Channel list: slot number out of range

 

 

The specified slot number is invalid. The channel number has the

 

 

form (@scc), where s is the slot number (100, 200, or 300) and cc is the

 

 

channel number. Example: CONF:VOLT:DC (@404)

 

112

Channel list: channel number out of range

 

 

The specified channel number is invalid for the module in the selected slot.

 

 

The channel number has the form (@scc), where s is the slot number

 

 

(100, 200, or 300) and cc is the channel number.

 

 

Example: ROUT:CLOSE (@134)

 

113

Channel list: empty scan list

 

 

Before you can initiate a scan, you must set up a scan list which includes

 

 

all configured multiplexer or digital channels in the instrument. Use the

 

 

MEASure?, CONFigure, or ROUTe:SCAN commands to set up your scan list.

 

201

Memory lost: stored state

 

 

This error is reported at power-on to indicate that a stored state has

 

 

become unusable. This error is most likely caused by a dead battery

 

 

(memory is battery-backed).Refer to the 34970A Service Guide to replace

 

 

the internal battery.

 

202

Memory lost: power-on state

 

 

 

This error is reported at power-on to indicate that the power-down state

 

 

of the instrument (normally recalled when power is turned on) has become

 

 

unusable. This error is most likely caused by a dead battery (memory is

6

 

battery-backed).Refer to the 34970A Service Guide to replace the

 

internal battery.

 

 

 

203

Memory lost: stored readings

 

 

This error is reported at power-on to indicate that readings stored in

 

 

memory from a previous scan have been lost. This error is most likely

 

 

caused by a dead battery (memory is battery-backed). Refer to the

 

 

34970A Service Guide to replace the internal battery.

 

204

Memory lost: time and date

 

 

This error is reported at power-on to indicate that the time and date

 

 

settings have been lost (they are reset to JAN 1, 1996 00:00:00).

 

 

This error is most likely caused by a dead battery (memory is

 

 

battery-backed).Refer to the 34970A Service Guide to replace the

 

 

internal battery.

 

309

Page 311
Image 311
Agilent Technologies 34970A manual Instrument Errors

34970A specifications

Agilent Technologies 34970A is a versatile data acquisition and measurement system that has gained popularity in various industries, including research and development, manufacturing, and education. The primary purpose of the 34970A is to offer an efficient solution for data collection, monitoring, and analysis, which can significantly enhance productivity and accuracy in testing applications.

One of the main features of the 34970A is its modular design, allowing users to customize the system according to their specific measurement needs. The mainframe can accommodate up to three plug-in measurement modules, which can include various types of measurements such as analog, digital, temperature, and frequency. This modularity provides flexibility for users to tailor the system to their requirements, making it highly adaptable to different applications.

Another notable characteristic of the Agilent 34970A is its impressive channel configuration. The system can support up to 120 measurement channels when fully configured, enabling extensive data acquisition without the need for multiple devices. This capability is essential for applications requiring simultaneous monitoring of multiple parameters or locations.

The 34970A utilizes advanced signal processing technologies to ensure high-accuracy measurements. The internal 6.5-digit resolution multimeter provides precise readings, while the device also supports various input types, such as thermocouples and RTDs for temperature measurements. The ability to perform mathematical functions, such as summation and averaging, on the acquired data further enhances its usability.

For data storage and management, the Agilent 34970A features built-in memory for storing up to 20,000 readings. Users can also easily transfer data to a PC through the RS-232 interface or GPIB (IEEE 488) for further analysis and reporting. Additionally, it offers the capability to program automatic data logging and scheduling of measurements, which streamlines the testing process and reduces manual intervention.

In terms of user interface, the 34970A is equipped with a large graphical display that provides clear visualization of measurement data and easy navigation through settings and options. The menu-driven interface makes it accessible for users of all skill levels, simplifying the setup and operation of the device.

Overall, the Agilent Technologies 34970A stands out for its modularity, high channel capacity, advanced measurement technologies, and user-friendly interface. As a reliable and efficient data acquisition system, it is well-suited for a wide range of applications, making it an essential tool for engineers and technicians looking to enhance measurement precision and efficiency.