Totalizer The multifunction module has a 26-bit totalizer which can
count pulses at a 100 kHz rate. You can manually read the totalizer
count or you can configure a scan to read the count.
You can configure the totalizer to count on the rising edge or falling
edge of the input signal.
The maximum count is 67,108,863 (226- 1). The count rolls over to “0”
after reaching the maximum allowed value.
You can configure the totalizer to read without affecting the count or
reset the count to zero without losing any counts.
+IN
Gate
-IN
Gate
Totalize
26 Bits Channel 03
Chapter 3 System Overview
Measurement Input