•Remote Interface Operation: Specify the resolution in the same units
as the measurement function, not in number of digits. For example,
if the function is dc volts, specify the resolution in volts. Fo r frequency,
specify the resolution in hertz.
You can select the resolution using parameters in the MEASure? and
CONFigure commands. For example, the following statement selects
the 10 Vdc range with 41⁄2 digits of resolution on channel 301.
CONF:VOLT:DC 10,0.001,(@301)
The following statement selects the 1 A range with 61⁄2 digits of
resolution on channel 221.
MEAS:CURR:AC? 1,1E-6,(@221)
You can also select the resolution usi ng the SENSe commands.
For example, the following statement specifies a 4-wire ohms
measurement with 100Ω of resolution on channel 103.
SENS:FRES:RES 100,(@103)
Chapter 4 Features and Functions
General Measurement Configuration
102