Remote Interface Operation: Specify the resolution in the same units
as the measurement function, not in number of digits. For example,
if the function is dc volts, specify the resolution in volts. Fo r frequency,
specify the resolution in hertz.
You can select the resolution using parameters in the MEASure? and
CONFigure commands. For example, the following statement selects
the 10 Vdc range with 412 digits of resolution on channel 301.
CONF:VOLT:DC 10,0.001,(@301)
The following statement selects the 1 A range with 612 digits of
resolution on channel 221.
MEAS:CURR:AC? 1,1E-6,(@221)
You can also select the resolution usi ng the SENSe commands.
For example, the following statement specifies a 4-wire ohms
measurement with 100 of resolution on channel 103.
SENS:FRES:RES 100,(@103)