SYSTem:ERRor?
Query the instrument’s error queue. A record of up to 10 errors is s tored
in the instrument’s error queue. Errors are retrieved in first-in-first-out
(FIFO) order. The first error returned is the first error that was stored.
When you have read all errors from the queue, the ERROR annunciator
turns off and the errors are cleared. The er ror queue is cleared by the
*CLS (clear status) command or when power is cycled. The errors are
also cleared when you read the queue.
See chapter 6 for a complete listing of the 34970A error messages.
SYSTem:ALARm?
Read the alarm data from the alarm queue (one alarm event is read and
cleared each time this command is executed). The following is an example
of an alarm stored in the alarm queue (if no alarm data is in the queue,
the command returns “0” for each field).
SYSTem:VERSion?
Query the instrument to determine the present SCPI version. Returns a
string in the form “YYYY.V”, where “YYYY” represents the year of the
version, and “V” represents a version number for that year (e.g., 1994.0).
*TST?
Perform a complete self-test of the instrument. Returns “+0” if the
self-test is successful, or “+1” if the test fails.
1 Reading with Units (31.009 °C)
2 Date (May 1, 1997)
3 Time (2:39:40.058 PM)
4 Channel Number
5 Limit Threshold Crossed (0 = No Alarm,
1 = LO, 2 = HI)
6 Alarm Number Reported (1, 2, 3, or 4)
Chapter 5 Remote Interface Reference
System-Related Commands
268