CONFigure:VOLTage:DC
CONFigure:VOLTage:AC
[{<range>|AUTO|MIN|MAX|DEF}
[,<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
Configure the specified channels for dc or ac voltage measurements but
do not initiate the scan. Note that this command also redefines the scan list.
For ac measurements, the resolution is actually fixed at 612 digits;
the resolution parameter only affects the number of digits shown on the
front-panel.
CONFigure:RESistance
CONFigure:FRESistance
[{<range>|AUTO|MIN|MAX|DEF}
[,<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
Configure the specified channels for 2-wire or 4-wire measurements but
do not initiate the scan. Note that this co mm an d also redefines the scan list.
For 4-wire measurements (FRES), the instrument automatically pairs
channel n with channel n+10 (34901A) or n+8 (34902A) to provide the
source and sense connections. For example, make the source connections
to the HI and LO terminals on channel 2 and the sense connections to
the HI and LO terminals on channel 12. Specify the paired channel in
the lower bank (source) as the scan_list channel.
CONFigure:CURRent:DC
CONFigure:CURRent:AC
[{<range>|AUTO|MIN|MAX|DEF}
[,<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
Note: Current measurements are allowed only on channels 21 an d 22
on the 34901A multiplexer module.
Configure the specified channels for dc or ac current measurements but
do not initiate the scan. Note that this command also redefines the scan list.
For ac measurements, the resolution is actually fixed at 612 digits;
the resolution parameter only affects the number of digits shown on
the front-panel.
Chapter 5 Remote Interface Reference
The MEASure? and CONFigure Commands
212