Chapter 6 Error Messages

Instrument Errors

501

I/O processor: isolator framing error

502

I/O processor: isolator overrun error

511

Communications: RS-232 framing error

512

Communications: RS-232 overrun error

513

Communications: RS-232 parity error

514

RS-232 only: unable to execute using HP-IB

 

There are three commands which are allowed only with the RS-232

 

interface: SYSTem:LOCal, SYSTem:REMote, and SYSTem:RWLock.

521

Communications: input buffer overflow

522

Communications: output buffer overflow

532

Not able to achieve requested resolution

 

The instrument cannot achieve the requested measurement resolution.

 

You may have specified an invalid resolution in the CONFigure or

 

MEASure? command.

540

Not able to null channel in overload

 

The instrument cannot store an overload reading (9.90000000E+37)

 

as the offset for Mx+B scaling using a null measurement.

550

Not able to execute command in local mode

 

The instrument has received a READ? or MEASure? command while in

 

the local mode.

6

313

Page 315
Image 315
Agilent Technologies 34970A manual Instrument Errors 501, 502, 511, 512, 513, 514, 521, 522, 532, 540, 550

34970A specifications

Agilent Technologies 34970A is a versatile data acquisition and measurement system that has gained popularity in various industries, including research and development, manufacturing, and education. The primary purpose of the 34970A is to offer an efficient solution for data collection, monitoring, and analysis, which can significantly enhance productivity and accuracy in testing applications.

One of the main features of the 34970A is its modular design, allowing users to customize the system according to their specific measurement needs. The mainframe can accommodate up to three plug-in measurement modules, which can include various types of measurements such as analog, digital, temperature, and frequency. This modularity provides flexibility for users to tailor the system to their requirements, making it highly adaptable to different applications.

Another notable characteristic of the Agilent 34970A is its impressive channel configuration. The system can support up to 120 measurement channels when fully configured, enabling extensive data acquisition without the need for multiple devices. This capability is essential for applications requiring simultaneous monitoring of multiple parameters or locations.

The 34970A utilizes advanced signal processing technologies to ensure high-accuracy measurements. The internal 6.5-digit resolution multimeter provides precise readings, while the device also supports various input types, such as thermocouples and RTDs for temperature measurements. The ability to perform mathematical functions, such as summation and averaging, on the acquired data further enhances its usability.

For data storage and management, the Agilent 34970A features built-in memory for storing up to 20,000 readings. Users can also easily transfer data to a PC through the RS-232 interface or GPIB (IEEE 488) for further analysis and reporting. Additionally, it offers the capability to program automatic data logging and scheduling of measurements, which streamlines the testing process and reduces manual intervention.

In terms of user interface, the 34970A is equipped with a large graphical display that provides clear visualization of measurement data and easy navigation through settings and options. The menu-driven interface makes it accessible for users of all skill levels, simplifying the setup and operation of the device.

Overall, the Agilent Technologies 34970A stands out for its modularity, high channel capacity, advanced measurement technologies, and user-friendly interface. As a reliable and efficient data acquisition system, it is well-suited for a wide range of applications, making it an essential tool for engineers and technicians looking to enhance measurement precision and efficiency.