Automatic Channel Delays

If you do not specify a channel delay, the instrument selects a delay

for you. The delay is determined by function, range, integration time,

and ac filter setting as shown below.

DC Voltage, Thermocouple, DC Current (for all ranges):

Integration Time Channel Delay
PLC > 1
PLC 1 2.0 ms
1.0 ms

Resistance, RTD, Thermistor (2- and 4-wire):

Range Channel Delay
(For PLC > 1) Range Channel Delay
(For PLC 1)
100
1 k
10 k
100 k
1 M
10 M
100 M
2.0 ms
2.0 ms
2.0 ms
25 ms
30 ms
200 ms
200 ms
100
1 k
10 k
100 k
1 M
10 M
100 M
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms

AC Voltage, AC Current (for all ranges):

AC Filter Channel Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
7.0 sec
1.0 sec
120 ms

Frequency, Period:

AC Filter Channel Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
0.6 sec
0.3 sec
0.1 sec

Digital Input, Totalize:

Channel Delay
0 sec
4

Chapter 4 Features and Functions

Scanning

89