E2 – Elevation Out Of Range
Step 1. Calibrate Elevation
Calibrate the elevation as outlined in the System Calibration Procedure in below step.
Error 2
Enter engineer mode of
“MANUAL
CALIBRATION”
screen
Check the “P13 ELV DIRECTION” screen whether set to “Yes”
Enter the “P4 ELEVATION
MINIMUM” screen
whether parameter set is
30 2
Enter the “P5
ELEVATION
MAXIMUM ” screen
whether parameter set is
240 5
Step 2. CORRECTIVE ACTION
Set to “Yes” | AUTO- | |
Calibration | ||
|
AUTO-
Calibration
AUTO-
Calibration
1.Check the failed system with a known good console cable. If fixed then repair/replace the console cable.
2.Check the failed system with a known good elevation motor. If problem fixed then repair/replace the elevation motor.
3.Check the failed system with a known good interface board. If problem fixed then repair/replace the interface board (MCB).
4.Check the failed system with a known good console board with current software. If the problem is fixed then repair/replace the console.
33