3M 747 manual Open-Collector Output, Shoe and Wrist Strap ALL Setting

Page 18

Open-Collector Output

When the SIGNAL OUT switch is set to ALL:

The output transistor switches on when the results of the shoe and wrist strap test are OK. When making alternating shoe/wrist strap tests, the type of test to be conducted next is indicated by the blinking symbol on the LCD and the blinking LED.

When the SIGNAL OUT switch is set to WRIST:

The output transistor is on when the result of a wrist strap test is OK.

When the SIGNAL OUT switch is set to SHOES:

The output transistor is on when the result of a shoe test is OK.

Shoe and Wrist Strap (ALL) Setting

If you attempt to test while the wrist strap LED is blinking and without having the wrist strap connected to the Shoes and Wrist Strap Tester 747, “Err.2” appears on the LCD. In this case, connect the wrist strap cable to the front panel connector and test again.

Conducting a wrist strap test followed immediately by a shoe test allows the shoe test to be performed while the wrist strap remains connected to the front panel when testing one shoe only.

Important Note:

Using the 3MSoftware 747DLS with the 3MShoes and Wrist Strap Tester 747 requires the following action: When testing a wrist strap and two shoes, the wrist strap ground cord must be disconnected from the Shoes and Wrist Strap Tester 747 after testing the wrist strap to perform the shoes test.

If more than ten seconds elapse between a shoe test and a wrist strap test, the automatic power-off function clears the last test results. In this case, repeat the last test.

17

Image 18
Contents 3M Shoes and Wrist Strap Tester Table of Contents Introduction InspectionExplanation of Safety Label Signal Words & Symbols Intended UsePage Overview Product OutlineLCD Indicators Side Panels Back SidePreparation Before Testing Connecting the ESD Shoe Testing PlateShoes Power Preparation Installing and Replacing the Batteries Power Saver Function Turning Power On and OffTurning Power Off AC Adapter ConnectionTesting Wall MountingALL setting is described in .1, Open Collector Output ESD Shoe Testing with the 3M Shoes and Wrist Strap TesterImportant Note Use single-conductor wrist straps only External Output Function Open Collector OutputTest Result Output Settings Test Result Output TerminalInternal Circuit Shoe and Wrist Strap ALL Setting Open-Collector OutputSpecifications RS-232C CommunicationsPC Connection Cable ConnectionsPC System Requirements Measurement Section Measurement Resistance Range 0 to 200.0 MΩMaintenance & Service Battery Replacement IndicatorError Messages LCD can display the following error messagesCleaning Testing ConsiderationsOperational Verification Test Testing Wrist Strap Ranges Equipment RequiredTest Set-Up Testing Wrist Low Limit Refer to Wrist Test Table Testing Wrist High Limits Refer to Wrist Test TableWrist Test Table Testing Shoe Ranges Testing Shoes Low Limits Refer to Shoes Test TableWeee Statement Regulatory InformationShoe Test Table CULus StatementFCC Electronic Solutions Division Static Control Products
Related manuals