3M 747 manual Error Messages, LCD can display the following error messages

Page 23

Important Notes:

Remove the batteries before storage to prevent possible corrosion cause by battery leakage if the Shoes and Wrist Strap Tester 747 will not be used for a long period of time.

When using the AC adapter, remove the batteries from the Shoes and Wrist Strap Tester 747 to prevent corrosion due to possible electrolyte leakage.

When replacing the batteries, do not install old batteries with new ones, and do not mix different types of batteries. Check the battery polarity carefully when inserting the batteries.

Keep used batteries out of the reach of children. Dispose of used batteries according to their type in the prescribed manner and in the proper location.

When the battery symbol () appears on the LCD, replace the batteries. Refer to Section 2.3.1 “Installing and Replacing the Batteries.”

6.2 Error Messages

The LCD can display the following error messages:

Error Message

Err.1

Meaning

Finger removed from touch panel during measurement

Solution

Repeat the test

Err.2

Actual test type does not match

 

specified test type

O.F.

Measured value exceeded 200.0 M

 

(Overflow)

Check test type (location), and repeat the test

Check test type and connections, and repeat the test

6.3 Before Returning the 3MShoes and Wrist Strap Tester 747 for Service

CAUTION

To reduce the risks associated with hazardous voltage, which, if not avoided, may result in minor or moderate injury and/or property damage:

• Not intended to be serviced by the user. No used serviceable parts.

22

Image 23
Contents 3M Shoes and Wrist Strap Tester Table of Contents Inspection IntroductionIntended Use Explanation of Safety Label Signal Words & SymbolsPage Product Outline OverviewLCD Indicators Back Side Side PanelsConnecting the ESD Shoe Testing Plate Preparation Before TestingShoes Power Preparation Installing and Replacing the Batteries AC Adapter Connection Turning Power On and OffTurning Power Off Power Saver FunctionWall Mounting TestingESD Shoe Testing with the 3M Shoes and Wrist Strap Tester ALL setting is described in .1, Open Collector OutputImportant Note Use single-conductor wrist straps only Open Collector Output External Output FunctionInternal Circuit Test Result Output SettingsTest Result Output Terminal Open-Collector Output Shoe and Wrist Strap ALL SettingRS-232C Communications SpecificationsPC System Requirements PC ConnectionCable Connections Measurement Resistance Range 0 to 200.0 MΩ Measurement SectionBattery Replacement Indicator Maintenance & ServiceLCD can display the following error messages Error MessagesOperational Verification Test CleaningTesting Considerations Test Set-Up Testing Wrist Strap RangesEquipment Required Wrist Test Table Testing Wrist Low Limit Refer to Wrist Test TableTesting Wrist High Limits Refer to Wrist Test Table Testing Shoes Low Limits Refer to Shoes Test Table Testing Shoe RangesCULus Statement Regulatory InformationShoe Test Table Weee StatementFCC Electronic Solutions Division Static Control Products