3M 747 manual Equipment Required, Test Set-Up, Testing Wrist Strap Ranges

Page 25

Equipment Required

Resistance Substitution Box (RSB) – 630 KΩ to 103 MΩ (minimum range) ± 1% or discrete resistors ± 1%, using resistor values as indicated in the tabled below.

Lead wires – As required to connect between RSB and the 3MShoes and Wrist Strap Tester 747.

3MCopper Foil Tape with conductive adhesive or equivalent (4" L x 1" W recommended).

Test Set-Up

Apply a strip of copper foil tape to the unit’s “Touch Panel” as shown in Fig. 1, leaving a short tab for the purpose of attachment of a test lead. To prevent the copper foil tape from curling during removal of the release liner, peel the release liner from the copper foil tape.

Testing Wrist Strap Ranges

Connect the reference resistor to the “Wrist Strap Connector” and the tab of the Copper Foil Tape on the Shoes and Wrist Strap Tester 747 as shown in Figure 1.

3MCOPPER FOIL TAPE

3MSHOES AND WRIST

RSB

STRAP SENSOR

Figure 1: Wrist Strap Test Set-up

24

Image 25
Contents 3M Shoes and Wrist Strap Tester Table of Contents Inspection IntroductionIntended Use Explanation of Safety Label Signal Words & SymbolsPage Product Outline OverviewLCD Indicators Back Side Side PanelsConnecting the ESD Shoe Testing Plate Preparation Before TestingShoes Power Preparation Installing and Replacing the Batteries Turning Power Off Turning Power On and OffPower Saver Function AC Adapter ConnectionWall Mounting TestingESD Shoe Testing with the 3M Shoes and Wrist Strap Tester ALL setting is described in .1, Open Collector OutputImportant Note Use single-conductor wrist straps only Open Collector Output External Output FunctionTest Result Output Terminal Test Result Output SettingsInternal Circuit Open-Collector Output Shoe and Wrist Strap ALL SettingRS-232C Communications SpecificationsCable Connections PC ConnectionPC System Requirements Measurement Resistance Range 0 to 200.0 MΩ Measurement SectionBattery Replacement Indicator Maintenance & ServiceLCD can display the following error messages Error MessagesTesting Considerations CleaningOperational Verification Test Equipment Required Testing Wrist Strap RangesTest Set-Up Testing Wrist High Limits Refer to Wrist Test Table Testing Wrist Low Limit Refer to Wrist Test TableWrist Test Table Testing Shoes Low Limits Refer to Shoes Test Table Testing Shoe RangesShoe Test Table Regulatory InformationWeee Statement CULus StatementFCC Electronic Solutions Division Static Control Products