3M 747 manual Operational Verification Test, Cleaning, Testing Considerations

Page 24

Problem

Items to check

Measurement does not start

Are the batteries depleted?

when pressing the touch panel

Is the AC adapter connected? (Sec. 2.3.2)

 

Is the power switch on? (Sec. 2.4)

Abnormal measurement values

Is the shoe testing plate connected? (Sec. 2.1)

Abnormal test results

Are the threshold resistance switch selections correct? (Sec. 2.2)

Test results not available at

Is the SIGNAL OUT switch selection correct? (Sec. 4.1)

OUTPUT terminal

 

RS-232C communication

Is the 3MShoes and Wrist Strap Tester 747 running on the AC adapter?

inoperable

(Sec. 2.3.2)

 

Are the serial communications settings correct? (Sec. 4.2)

 

Is a crossed cable being used for the connection? (Sec. 4.2)

6.4 Cleaning

Gently wipe dirt from the surface of the unit with a soft cloth moistened with a small amount of water or mild detergent.

7.0 Operational Verification Test

CAUTION

To reduce the risks associated with hazardous voltage, which, if not avoided, may result in minor or moderate injury and/or property damage:

• Not intended to be serviced by the user. No user serviceable parts.

To reduce the risks associated with electrostatic discharge (ESD), which, if not avoided, may result in property damage to electronic components or assemblies being handled:

Perform operational verification test to ensure proper operations of the 3MShoes and Wrist Strap tester 747 as required.

Testing Considerations

This test procedure is used to verify that the 3MShoes and Wrist Strap Tester 747 functions within the specifications as stated in Chapter 5 of this instruction manual. It is recommended that you become familiar with the operation of the Shoes and Wrist Strap Tester 747 outlined in this manual, before beginning testing.

It is recommended that this procedure be performed annually, However, if the unit is exposed to a harsh environment or has been abused in some manner, it may require performing this test procedure more frequently.

23

Image 24
Contents 3M Shoes and Wrist Strap Tester Table of Contents Introduction InspectionExplanation of Safety Label Signal Words & Symbols Intended UsePage Overview Product OutlineLCD Indicators Side Panels Back SidePreparation Before Testing Connecting the ESD Shoe Testing PlateShoes Power Preparation Installing and Replacing the Batteries Turning Power On and Off Turning Power OffPower Saver Function AC Adapter ConnectionTesting Wall MountingALL setting is described in .1, Open Collector Output ESD Shoe Testing with the 3M Shoes and Wrist Strap TesterImportant Note Use single-conductor wrist straps only External Output Function Open Collector OutputTest Result Output Settings Test Result Output TerminalInternal Circuit Shoe and Wrist Strap ALL Setting Open-Collector OutputSpecifications RS-232C CommunicationsPC Connection Cable ConnectionsPC System Requirements Measurement Section Measurement Resistance Range 0 to 200.0 MΩMaintenance & Service Battery Replacement IndicatorError Messages LCD can display the following error messagesCleaning Testing ConsiderationsOperational Verification Test Testing Wrist Strap Ranges Equipment RequiredTest Set-Up Testing Wrist Low Limit Refer to Wrist Test Table Testing Wrist High Limits Refer to Wrist Test TableWrist Test Table Testing Shoe Ranges Testing Shoes Low Limits Refer to Shoes Test TableRegulatory Information Shoe Test TableWeee Statement CULus StatementFCC Electronic Solutions Division Static Control Products