Teledyne 465M instruction manual Span Calibration

Page 40

NOTE

While measuring zero air, a certain amount of noise or “dithering” of the concentration about the zero point will occur and is normal. This noise is typically 1-3 PPM in magnitude.

7.2.3.5. Span Calibration

The Span Calibration also allows the instrument to calibrate its internal ozone offset factor. The Span Calibration however, is done with some controlled concentration gas connected to the “Sample In” port of the M465M. It is recommended that a span gas is concentrated to around 80% of the operating range. Set the SPAN VALUE to the measured concentration of your calibration gas. This normally takes 10-15 minutes.

NOTE

A Span Calibration should only be performed with a stable source of Ozone and a reference monitor measuring the same gas supply.

If you are unsure regarding the suitability of a particular source of calibration gas, contact Customer Service at Teledyne API for assistance (see Section 12.)

See Figure 7-4shows the Span Cal menu. After the instrument has stabilized on the source of span gas, navigate to the Span Cal menu (See Figure 7-4) and Press ENT. Next enter the Span Target concentration (the actual concentration of ozone being supplied to the monitor) and press ENT. The display will next show a confirmation menu, ENT TO SPAN. Press ENT to perform the Span Calibration, or CFG to abort back to the start. If the calibration is successful, the display will return to the concentration menu and the monitor reading should change adjust to read very close to the target value. If the calibration cannot be performed, an OUT OF RANGE error will be displayed. Press ENT to confirm and the display will return to the start of the Span Cal menu.

If the OUT OF RANGE error occurs, it means the Span Cal cannot be performed because it would result in an out of range slope value for the monitor. This means that either the sensor in the monitor is malfunctioning, causing improper readings, or the actual ozone concentration being supplied to the monitor is different than the target value being entered.

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 38

Image 40
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsSafety ApprovalsEMC Other Type CertificationsPrinciple of Operation 3  Page Mechanical Installation for Nema 4X Enclosure Getting StartedVentilation Clearance Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Electrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status Outputs Status Name Normal Operating Output # StateHi-Current Relay Outputs Digital Serial InterfacesRelay # Function Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Optional Hardware and Software Low Flow AlarmRack Mount Slides This page is Intentionally Left Blank Operation Front Panel OverviewOzone Concentration Display Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStatus LED’s Status LED Normal StateMenu Navigation Basics Front Panel MenusEditing Numeric Values Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeMeaning Default Value SETUP-VARS MenuRange Config Menu SETUP-ALARM MenuProperty Meaning Default Range Value This page is Intentionally Left Blank Digital Communications Changing COM port Hardware ConfigurationDigital Communications Overview RS232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-232 ConfigurationProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEEthernet Configuration Hardware Setup for Configuring the Ethernet ModuleVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Instrument Setup and Configuration Registers Read / Write Modbus Register MapsReal-Time Concentration and Status Registers Read Only Decimal Star Address Description Register Type Bit BitsMaintenance Schedule MaintenanceMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement UV Lamp Replacement UV Lamp Adjustment1150 mV Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Hardware Setup Boot Loader OperationBoot Loader Software Setup Upload Firmware Page Pneumatic Diagram TroubleshootingReference Drawings Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Pneumatic Error Sensor OKInvalid Reading Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version