Figure 7-4: Span Cal Menu

7.2.4.VIEW Menu

The VIEW menu allows the operator to view various measurement parameters in real- time. This can be useful for diagnosing various instrument or system problems.

NOTE

These values are updated in real-time as they are repeatedly scrolled on the display.

Table 7-5: VIEW Menu Parameters

 

Parameter

 

Meaning

 

Units

 

Normal Range*

 

 

 

 

 

 

 

 

 

MEAS

 

UV detector reading, measure cycle

 

mV

 

250 – 1230

 

 

 

 

 

 

 

 

 

REF

 

UV detector reading, reference

 

mV

 

250 – 1230

 

 

cycle

 

 

 

 

 

 

 

 

 

 

PRESSURE

 

Sample pressure

 

psia

 

9.0 – 18.0

 

 

 

 

 

 

 

 

 

LAMP

 

UV lamp temperature

 

ºC

 

50.0 – 52.0

 

TEMP

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

Sample temperature

 

ºC

 

5.0 –50.0

 

TEMP

 

 

 

 

 

 

 

 

 

 

 

FLOW

 

Sample Gas Flow Rate**

ON/OFF

 

ON

 

 

 

 

 

 

 

 

*After initial 20 minute warm-up period.

 

 

 

 

 

** Optional Hardware.

 

 

 

 

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 39

Image 41
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank General Safety Hazard Safety MessagesMise EN Garde Consignes DE SécuritéWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Preface IntroductionThis page is Intentionally Left Blank Specifications Specifications and Agency ApprovalsApprovals SafetyEMC Other Type Certifications3   Principle of OperationPage Getting Started Mechanical Installation for Nema 4X EnclosureVentilation Clearance Area Minimum required clearanceM465M Mounting Hole Locations and Dimensions Mechanical Installation 19 Rack VersionAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationOzone Inlet Connection Pneumatic Connections, Nema ConfigurationRear Panel Electrical I/O Connections Exhaust ConnectionNema Electrical I/O Connections Analog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status Name Normal Operating Output # State Status OutputsDigital Serial Interfaces Hi-Current Relay OutputsRelay # Function Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Rack Mount Slides Low Flow AlarmOptional Hardware and Software This page is Intentionally Left Blank Front Panel Overview OperationOzone Concentration Display Menu Navigation KeypadAlarm Status LED Color Alarm State Concentration Alarm Status/ResetStatus LED’s Status LED Normal StateEditing Numeric Values Front Panel MenusMenu Navigation Basics Front Panel Menu Diagram Front Panel Menu Diagram Point 5V Output MA Output Diag MenuZero Calibration Signal Name Rear Panel Label FunctionSpan Calibration Parameter Meaning Units Normal Range View MenuRange SETUP-VARS MenuMeaning Default Value Property Meaning Default Range Value SETUP-ALARM MenuConfig Menu This page is Intentionally Left Blank Digital Communications Overview Changing COM port Hardware ConfigurationDigital Communications RS232 Configuration Property Value RS-232 ConfigurationFunction DB-9 General I/O Connector Pin SW2 DTE SW2 DCE Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationHardware Setup for Configuring the Ethernet Module Ethernet ConfigurationVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModePage Modbus Register Maps Instrument Setup and Configuration Registers Read / WriteReal-Time Concentration and Status Registers Read Only Decimal Star Address Description Register Type Bit BitsMaintenance Item Recommended Interval Section MaintenanceMaintenance Schedule Instrument Layout Instrument Layout, Rack Mount ConfigurationInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement 1150 mV UV Lamp AdjustmentUV Lamp Replacement Auto-Zero Valve Replacement Sample Pump ReplacementConfiguring the Analog Output Zero Cleaning Exterior Surfaces of the M465MBoot Loader Software Setup Boot Loader OperationBoot Loader Hardware Setup Upload Firmware Page Reference Drawings TroubleshootingPneumatic Diagram Pneumatic Diagram Nema Version with Sample Conditioner Interconnect DiagramInterconnect Diagram Status LED/Output Definitions Sensor OK Pneumatic ErrorInvalid Reading Check LampPressure Too High Factory Calibration CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank 13. M465M Spare Parts Spare Parts List for Rack Mount VersionSpare Parts List for Nema Version