7.2.3.DIAG Menu

The DIAG menu (See Figure 7-2and Figure 7-3)contains functions that are useful for testing and configuring external equipment that may be connected to the rear panel I/O of the instrument.

7.2.3.1. Analog Step-Test

When placed in this mode, the instrument will automatically step the analog output through 5 points from 0 to 100 % and display the current value on the front panel. The operator can suspend the automatic cycling and manually control the stepping by pressing either one of the UP / DOWN arrow buttons. To resume automatic cycling, exit the function by pressing CFG and re-enter by pressing ENT.

The 5 points will correspond to the following analog output values depending on whether the output is configured for 0-5V or 4-20mA:

Table 7-3: Analog Step Test Values

Point

0-5V Output

4-20 mA Output

0%

0.00 V ± 0.02

4 mA ± 0.02

25%

1.25 V ± 0.02

8 mA ± 0.02

50%

2.50 V ± 0.02

12 mA ± 0.02

75%

3.75 V ± 0.02

16 mA ± 0.02

100%

5.00 V ± 0.02

20 mA ± 0.02

Press CFG to exit from the Analog Step-Test function.

7.2.3.2. Signal I/O

The Signal I/O menu allows the operator to manually control the various relay outputs and control inputs available on the rear panel of the M465M. This function is useful for testing or debugging external control systems (i.e. dataloggers or PLC’s) to which the instrument may be connected.

See Section 4.5 for additional details on making connections to these I/O signals on the rear panel.

See Table 7-4for additional details on entering the Signal I/O menu.

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 36

Image 38
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsEMC SafetyApprovals Other Type CertificationsPrinciple of Operation 3  Page Ventilation Clearance Mechanical Installation for Nema 4X EnclosureGetting Started Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Analog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status Outputs Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsDigital Serial Interfaces Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Rack Mount Slides Low Flow AlarmOptional Hardware and Software This page is Intentionally Left Blank Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadStatus LED’s Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Status LED Normal StateEditing Numeric Values Front Panel MenusMenu Navigation Basics Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeRange SETUP-VARS MenuMeaning Default Value Property Meaning Default Range Value SETUP-ALARM MenuConfig Menu This page is Intentionally Left Blank Digital Communications Overview Changing COM port Hardware ConfigurationDigital Communications RS232 Configuration Property Value RS-232 ConfigurationFunction DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Real-Time Concentration and Status Registers Read Only Instrument Setup and Configuration Registers Read / WriteModbus Register Maps Decimal Star Address Description Register Type Bit BitsMaintenance Item Recommended Interval Section MaintenanceMaintenance Schedule Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement 1150 mV UV Lamp AdjustmentUV Lamp Replacement Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Software Setup Boot Loader OperationBoot Loader Hardware Setup Upload Firmware Page Reference Drawings TroubleshootingPneumatic Diagram Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version