Teledyne 465M instruction manual Mechanical Installation 19 Rack Version

Page 18

Figure 4-1: M465M Mounting Hole Locations and Dimensions

4.2.Mechanical Installation (19” Rack Version)

The Rack Mount version of the M465M was designed to be operated as a bench-top monitor or to be mounted in a standard 19” RETMA instrumentation rack. For rack installations, the four mounting feet should be removed from the bottom of the monitor.

4.2.1.Ventilation Clearance

When installing the M465M be sure to leave sufficient ventilation clearance.

Table 4-2: M465M Ventilation Clearances, Rack Mount Version

Area

Minimum required clearance

Back of the monitor

3 inches

Sides of the monitor

1 inch

Above and below the monitor

1/2 inch

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 16

Image 18
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsEMC SafetyApprovals Other Type CertificationsPrinciple of Operation 3  Page Ventilation Clearance Mechanical Installation for Nema 4X EnclosureGetting Started Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Location of I/O Connectors in the Nema Configuration Electrical I/O ConnectionsAnalog Output Status Outputs Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsDigital Serial Interfaces Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Low Flow Alarm Optional Hardware and SoftwareRack Mount Slides This page is Intentionally Left Blank Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadStatus LED’s Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Status LED Normal StateFront Panel Menus Menu Navigation BasicsEditing Numeric Values Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeSETUP-VARS Menu Meaning Default ValueRange SETUP-ALARM Menu Config MenuProperty Meaning Default Range Value This page is Intentionally Left Blank Changing COM port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Real-Time Concentration and Status Registers Read Only Instrument Setup and Configuration Registers Read / WriteModbus Register Maps Decimal Star Address Description Register Type Bit BitsMaintenance Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement UV Lamp Adjustment UV Lamp Replacement1150 mV Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Operation Boot Loader Hardware SetupBoot Loader Software Setup Upload Firmware Page Troubleshooting Pneumatic DiagramReference Drawings Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version