8.4.RS-485 Configuration
RS-485 is commonly used for factory automation busses with a computer or PLC acting as a master. The communications protocol used for the RS-485 configuration is MODBUS RTU. For details on the MODBUS RTU specification, please see http://www.modbus.org/.
With the instrument configured for RS-485, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector. Table 8-3below details the pinouts of these two connectors:
Table 8-3: RS-485 Connector Details
| Function | DB-9 | General I/O | 
|   | Connector Pin | Connector | 
|   | (SW2 – DCE) |   | 
| RS485 - A | 2 | ‘T’ | 
| RS485 - B | 3 | ‘R’ | 
| Ground | 5 | ‘G’ | 
NOTE
For RS-485 configuration, the DTE-DCE switch should always be set to DCE. See Section 8.2 for details on changing this configuration.
The serial port setup for RS-485 configuration is shown in Table 8-4below:
Table 8-4: RS-485 Port Setup
| Property | Value | 
| Baud Rate | 57600 | 
| Data Bits | 8 | 
| Parity | None | 
| Stop Bits | 1 | 
| Flow Control | None | 